The flow has gone below 50ml/min. This can be caused by the probe being blocked, filter being contaminated, bent pneumatics, exhaust blockage or dirt and water ingress. If you see this fault, firstly check the probe and exhaust for blockages and replace the probe filter. If this issue continues, please reference the Tiger service manual, or talk to us here at ION Science.